CRYODEEP CryoDeep Technologies
Product Line 01

Cryogenic Level Measurement

A liquid helium level measurement solution using a superconducting filament as the sensing element. Meter and probe are supplied as one set, with an optional pulsed-current mode that reduces the helium loss caused by measurement. Suitable for superconducting magnets, particle accelerators, cryostats and liquid helium dewars.

DEEP-TWO dual-channel liquid helium level meter
Design Notes

Three choices made for long-term continuous monitoring

01

Pulsed-current measurement, less helium loss

Level measurement itself introduces heat into the liquid helium and causes evaporative loss. The pulsed-current mode lowers the average heating power while preserving accuracy, and so reduces the helium lost to measurement.

02

A four-wire measurement loop

The probe is wired in a four-wire configuration with 28 AWG copper leads, removing lead resistance by principle; combined with resistance measurement better than 0.1% in the meter, this yields a stable, repeatable level reading.

03

A slim G10 probe, 5.8 mm in diameter

The G10 outer tube is only 5.8 mm across, matching a wide range of liquid helium dewars and cryostats. Effective lengths of 200 / 400 mm and custom lengths are available.

Dual-Channel Liquid Helium Level Meter
Cryogenic level · DEEP-TWO-4K

Dual-Channel Liquid Helium Level Meter

Measures liquid helium level with an ultra-fine superconducting filament. Two independent display channels and a pulsed-current measurement mode that reduces helium loss.

Measurement accuracy
0.1%
Output current
10–100 mA, continuously adjustable
Display
4.3" touchscreen, 800×480
Interfaces
RS-232 · Ethernet · AUX I/O
Superconducting Level Probe
Cryogenic level · CRYO-4K-XXmm

Superconducting Level Probe

G10 outer tube in a slim 5.8 mm design, matching a wide range of liquid helium dewars and cryostats. Custom lengths available.

Effective length
200 / 400 mm / custom
Outer tube material / diameter
G10 / 5.8 mm
Excitation current
75 mA
Operating temperature range
4.2 K ± 0.2 K
Specifications

DEEP-TWO Level Meter System Specifications

Meter and probe are supplied as a set and must be purchased together. The values below are nominal factory figures; refer to the document supplied with the unit for actual parameters.

DEEP-TWO-4K front panel: CH1 level height 150 mm, CH2 helium inventory 75%
DEEP-TWO-4K front panel: CH1 level height 150 mm, CH2 helium inventory 75%
DEEP-TWO-4K rear panel: Ethernet, RS-232, AUX I/O 1/2 and two probe connectors; 100–240 Vac mains input
DEEP-TWO-4K rear panel: Ethernet, RS-232, AUX I/O and two probe connectors; 100–240 Vac input
CRYO-4K superconducting level probe: G10 outer tube, 5.8 mm diameter
CRYO-4K superconducting level probe: G10 outer tube, 5.8 mm diameter, four-wire 28 AWG leads

Level Meter

DEEP-TWO-4K
Display channels2, independently displayed
Output current10–100 mA continuously adjustable; control accuracy better than 1 mA
Measurement accuracy0.1% (filament resistance measurement better than 0.1%)
Displayed informationPage height and helium inventory (one-touch toggle)
Display size / resolution4.3" / 800×480
Measurement modePulsed current mode
Alarm functionLow-level alarm, can trigger an external alarm device
InterfacesRS-232, Ethernet, AUX I/O, DB9 level probe connector
Mains input100–240 Vac / 50–60 Hz, max input power 200 W
CommunicationTCP server mode, fixed local listening port 8080
Dimensions430 × 300 × 88.9 mm (2U chassis height)

Superconducting Level Probe

CRYO-4K-XXmm
Effective length200 mm / 400 mm / custom
Overall lengthEffective length + 3 mm (3 mm clearance below)
Outer tube materialG10
Outer tube diameter5.8 mm
Excitation current75 mA (recommended, configurable)
Voltage40 V
LeadsFour-wire, 28 AWG copper
Resistivity80.9 Ω/m @20 K
97.5 Ω/m @300 K
Liquid helium temperature range4.2 K ± 0.2 K
Nominal initial resistance32.7 Ω (CRYO-4K-200; recommended 90 mA / 3 s preheat)
Typical Configurations

Three common setups

A

Meter + single probe

Level monitoring for a single liquid helium dewar. The meter's second channel can be left spare for a future second dewar or a backup probe.

B

Meter + two probes

Simultaneous monitoring of two independent vessels in the same system (for example a magnet dewar and a cold box), displayed side by side on one panel.

C

Integration with a host system

Connect over Ethernet or RS-232 to a laboratory control system. The meter runs as a local TCP server for unified logging and alarm interlocking.

Not sure which probe length to choose?

Send us your dewar inner dimensions and mounting interface, and we will recommend the probe length and excitation current.